Design and evaluation of a reflectance diffuse optical tomography system


Kazanci H. Ö., Mercan T., CANPOLAT M.

OPTICAL AND QUANTUM ELECTRONICS, cilt.47, sa.2, ss.257-265, 2015 (SCI-Expanded) identifier identifier

  • Yayın Türü: Makale / Tam Makale
  • Cilt numarası: 47 Sayı: 2
  • Basım Tarihi: 2015
  • Doi Numarası: 10.1007/s11082-014-9910-6
  • Dergi Adı: OPTICAL AND QUANTUM ELECTRONICS
  • Derginin Tarandığı İndeksler: Science Citation Index Expanded (SCI-EXPANDED), Scopus
  • Sayfa Sayıları: ss.257-265
  • Anahtar Kelimeler: Reflectance diffuse optical tomography, Optode calibrations, Imaging systems, Medical and biological imaging, BRAIN ACTIVATION, SPATIAL-RESOLUTION, CALIBRATION
  • Akdeniz Üniversitesi Adresli: Evet

Özet

We have designed a continuous wave, back-reflection diffuse optical tomography system; and developed a new practical calibration method including both optode efficiency and positional error corrections. System design, data acquisition and calibration protocols are described in detail. Monte Carlo (MC) simulations of photon distribution for tissue phantoms have been used to obtain the weight matrix to be used in the Rytov approximation to the photon diffusion equation. The system has been evaluated by acquiring data from a tissue phantom with a background scattering coefficient () of and absorption coefficient ( of . An inclusion made of 1 % Intralipid and indocyanine green with and was placed at a 2 cm depth from the tissue phantom surface. After calibration, the average value of the measurements over source-detector pairs at the same distance for each neighborhood was calculated. Perturbation data were obtained by subtracting the average data from the measurements with the same source-detector separation. In the reconstruction, weight matrixes obtained from MC simulation for to were used. The Depth Compensation Algorithm was used in the Tikhonov regularization to identify the location of the inclusion correctly in the reconstruction.